Determination of trace levels of dinitrophenolic compounds in environmental water samples using hollow fiber supported liquid membrane extraction and high performance liquid chromatography
A hollow fiber supported liquid membrane extraction method for the liquid chromatographic determination of dinitrophenolic compounds at ppt levels has been developed. Different variables affecting the extraction process, such as extraction time, shaking speed, acceptor pH, acceptor buffer concentration, salt content and humic acids have been studied. Enrichment factors up to 7000 times were obtain