Temperature changes within samples in heat conduction calorimeters
So-called isothermal calorimeters are often built on the heat conduction principle. This may sound as a contradiction, as heat conduction is a result of non-isothermal conditions, but most measurements with microcalorimeters and other such instruments are essentially isothermal as the temperature differences within the instruments are in the order of 10 mK. This paper presents simple methods to es
