Movement pattern of the Exeter femoral stem: A radiostereometric analysis of 22 primary hip arthroplasties followed for 5 years
Background The design of the Exeter stem may facilitate distal migration, but radiostereometric analysis (RSA) studies have been limited to 2 years of follow-up. Patients and methods We followed migration of the Exeter femoral stems in 22 primary hip arthroplasties for 5 years with RSA. Results All stems migrated distally and the median migration at 2 years was 1.34 mm, while at 5 years it was 1